Troubleshooting
No response or garbage data
- Confirm 19200 8N1 unless you have intentionally reprogrammed the sensor with
#BAUD. - Verify TX/RX are not swapped and grounds are common.
- Respect 3.3 V max on the FDO2 UART pins.
- Wait โฅ 1 s after power-up before the first command.
Timeout or ProtocolError
- Increase
SetMeasureTimeoutMs(default 300 ms;#MRAWcan be longer at high#SETMmodes on device). - Ensure no other task is reading the same UART concurrently without mutual exclusion.
- If CRC is enabled on the device, verify you are on a driver revision that strips
the ` :
` suffix (current `main` does).
DeviceError after #ERRO
Call LastDeviceErrorCode() and map E using Table 1 in the data sheet ยง4.4
(for example -22 UART receive, -26 unknown header).
Status S not 0 or 1
The data sheet requires checking S every sample. If MoxyStatusOkForOxygen
returns false, treat pOโ / temperature as invalid for safety-critical use until
the fault clears (see bit definitions in fdo2::moxy_status).
Need more help?
Open an issue with UART capture,
firmware R from #VERS, and software revision.