FDO2-G2 UART protocol
This driver follows PyroScience FDO2-G2, document version v5 (03/2026), §4 Communication Interface.
Electrical and framing
- 3.0 V UART levels (absolute maximum 3.3 V on IO).
- 8N1, no handshake.
- Default baud after power-up: 19200 (1200 … 115200 programmable via
#BAUD; not wrapped in this driver yet). - Allow ~1 s after power-up before commands.
- Host commands end with carriage return (
0x0D), or CR+LF (0x0D 0x0A). - Device responses end with CR only.
- If CRC is enabled (
#CRCE 1), each response ends with ` :` before CR. The driver **strips** that suffix before tokenizing.
Implemented commands
| Command | Response shape | Driver API |
|---|---|---|
#VERS |
#VERS D N R S |
ReadVersion() |
#IDNR |
#IDNR N (64-bit decimal) |
ReadUniqueId() |
#MOXY |
#MOXY O T S |
MeasureMoxy() |
#MRAW |
#MRAW O T S D I A P H |
MeasureMraw() |
#LOGO |
#LOGO |
FlashLogo() |
Numeric decoding
- O — pO₂ in 10⁻³ hPa →
p_o2_hpa = O × 10⁻³. - T — temperature in m°C →
temp_c = T × 10⁻³. - S — unsigned 32-bit status. Normal operation: S = 0 or 1 only.
- #MRAW extras: D in millidegrees; I, A in µV → mV with
×10⁻³; P in µbar → mbar with×10⁻⁶; H in m%RH →%RHwith×10⁻³.
Constants for individual status bits live in fdo2::moxy_status in fdo2_types.hpp.
Volume % O₂
When pressure at the sensing membrane matches the vent / back-side pressure
represented by P in #MRAW:
%O₂ = 100 × pO₂ [hPa] / p_air [hPa] (mbar ≈ hPa for this ratio per data sheet §2).
Use VolumePercentO2(p_o2_hpa, pressure_mbar).
Errors
Responses beginning with #ERRO followed by signed E (Table 1, §4.4). When
DriverResult::error == DeviceError, read Driver::LastDeviceErrorCode().
Out of scope in this driver
Flash-writing commands (#CALO, #CAHI, #CRCE, #SETM, #BCST, #BAUD,
#WRUM, …) are not exposed: they consume flash endurance (~20k cycles
total) and require stable power during programming (data sheet warning).